[Abstract]Objective:Toinvestigatetheefectof infantrespiratoryrateonchestCTradiationdoseandimagequalityinhigh pitch mode(HPM)andlowpitch mode(LPM).Methods:Eighty infantsunderwentchestCTandwere divided into HPM group(40 cases)andLPMgroup(40cases).Among the80cases,52cases hadhighrespiratoryrate(RR),28 hadlow RR. ThevolumeCTdoseindex(CTDIvol),doselengthproduct(DLP)andeffectivedose(ED),overscandistance,objective andsubjectiveimagequalityat diferentRRof thetwogroups werecompared.Results:EDvaluesof HPMand LPM groups were ( 3.31±0.28) ,(1.79±0.15) mSv,the overscan distances were( 6.54±5.94, ),(4.02±2.51) cm,and the differences werestatistically significant(both P lt;0.05).The image quality scores of HPM group were better than those of LPM group(motionartifactscoresof1.6and2.3points,subjectivescoresof3.6and3.1points,espetively),withsatisialy significant differences( P =0.002,0.O04).The patients at high RR in HPM group had less motionartifact and better subjective score of image quality compared with the patients at high RR in LPM group (both P lt;0.05),while thepatients at low RR had no statistical diferences in motionartifact orsubjective score between HPMgroup and LPM group (both Pgt;0.05 ) Conclusions:HPMcanbeusedtoevaluatethelungsof infantswith higherRR,andcanshowbeterimagequality.LPMcan beused forlungassessment of infantswith low RR,which canreduce radiationdoseand maintain good image quality. [Key words] High pitch mode;Low pitch mode;Infant;Radiation dosage;Tomopraghy,X-ray computed
胸部CT被用來(lái)評(píng)估多種嬰幼兒疾病,但不能自主屏氣患兒無(wú)法獲得高質(zhì)量的靜止胸部CT圖像。Lell等1]采用雙源CT系統(tǒng)中大螺距模式(highpitchmode,HPM)技術(shù),無(wú)需鎮(zhèn)靜或全身麻醉即可獲得高質(zhì)量的圖像,但增大螺距會(huì)增加患兒輻射劑量。雙源CT系統(tǒng)中小螺距模式(low pitch mode,LPM)器官-管電流調(diào)節(jié)劑量調(diào)制技術(shù)2,能實(shí)時(shí)調(diào)節(jié)X線(xiàn)管電流、減少輻射,但會(huì)增長(zhǎng)采集時(shí)間,使自由呼吸狀態(tài)下患兒的檢查產(chǎn)生較多運(yùn)動(dòng)偽影。
胸部圖像上膈肌周?chē)谓M織和靠近胸壁的前肺組織運(yùn)動(dòng)偽影更明顯,與呼吸運(yùn)動(dòng)明顯關(guān)聯(lián)[3]。本研究通過(guò)分析嬰幼兒自由呼吸狀態(tài)下HPM和LPM雙源CT的輻射劑量和圖像質(zhì)量,探討不同RR對(duì)2種模式雙源CT胸部掃描輻射劑量和圖像質(zhì)量的影響。
1資料與方法
1.1一般資料
選取2022年1—10月在我院行胸部CT檢查的患兒80例,男46例,女34例;年齡 1~36 個(gè)月,平均( 15.5±5.4) 個(gè)月。80例中,病毒性肺炎29例,細(xì)菌性肺炎13例,肺水腫22例,腦癱合呼吸道感染16例。隨機(jī)分為HPM組和LPM組,每組40例,分別行HPM、LPM胸部CT檢查。檢查前測(cè)量RR和體質(zhì)量。80例中,高RR( gt;25 次 /min 052例,低RR( ≤ 25次 /min )28例。本研究經(jīng)醫(yī)院倫理委員會(huì)批準(zhǔn)(批號(hào):20230108),所有患兒家屬均簽署知情同意書(shū)。
1.2 儀器與方法
采用Siemens第3代雙源CT(SOMATOMForce),在患兒熟睡狀態(tài)下掃描。HPM螺距3.0,F(xiàn)lashSpiral模式自適應(yīng)動(dòng)態(tài)準(zhǔn)直,掃描速度 404mm/s ;LPM螺距0.6,X-CARE模式一定范圍內(nèi)關(guān)閉管電流,降低人體淺表器官(如乳房、晶狀體、胸腺和甲狀腺)的輻射劑量,掃描速度 81mm/s 。其他掃描參數(shù):探測(cè)器準(zhǔn)直寬度 128×0.6mm ,自動(dòng)管電壓和自動(dòng)管電流技術(shù)(參考管電壓 120kV ,參考管電流 100mA ,管電壓70~100kV ,管電流 33-92mA ,機(jī)架旋轉(zhuǎn)速度 0.28s/r 視野 189mm×189mm ,層厚 3mm ,層距 2mm ,ADMIR重建模式,圖像重建使用卷積核 I30f □
1.3輻射劑量估算
記錄檢查時(shí)間、CT容積劑量指數(shù)(CTDIvol)和劑量長(zhǎng)度乘積(doselengchproduct,DLP)。有效輻射劑量(effective dose,ED)=DLP×k,k=0.039[4] 。過(guò)度掃描距離
DLP/CTDIvol-(結(jié)束檢查位置-開(kāi)始檢查位置),單位 cm 。
1.4 圖像質(zhì)量評(píng)價(jià)
1.4.1主觀(guān)評(píng)價(jià)由2位高級(jí)職稱(chēng)影像診斷醫(yī)師在同一臺(tái)PACS上對(duì)運(yùn)動(dòng)偽影及圖像質(zhì)量進(jìn)行定性評(píng)估,意見(jiàn)不一致時(shí)經(jīng)協(xié)商解決。CT檢查分別在膈肌、左心緣、上肺、中肺和下肺5個(gè)部位進(jìn)行定性分析。膈肌通過(guò)冠狀面重建進(jìn)行評(píng)估,左心緣邊界在軸位圖像和冠狀面重建圖像上進(jìn)行評(píng)估,而上肺、中肺和下肺僅在橫軸位圖像上評(píng)估,上肺評(píng)估在主動(dòng)脈弓上緣水平,中肺評(píng)估在主動(dòng)脈瓣水平,下肺評(píng)估在二尖瓣水平。出現(xiàn)解剖結(jié)構(gòu)重影或肺血管紋理模糊即為有運(yùn)動(dòng)偽影。參照Likert量表,采用4分制對(duì)每個(gè)位置的運(yùn)動(dòng)偽影進(jìn)行評(píng)分:3分,存在大量運(yùn)動(dòng)偽影影響診斷為重度;2分,運(yùn)動(dòng)偽影明顯但不干擾診斷為中度;1分,運(yùn)動(dòng)偽影極少為輕度;0分,無(wú)任何運(yùn)動(dòng)偽影。采用5分制對(duì)圖像質(zhì)量進(jìn)行評(píng)分:5分,圖像質(zhì)量?jī)?yōu),噪聲小,細(xì)節(jié)顯示清晰,完全滿(mǎn)足診斷;4分,圖像質(zhì)量良,噪聲較小,細(xì)節(jié)顯示清晰,可滿(mǎn)足診斷;3分,圖像質(zhì)量一般,細(xì)節(jié)顯示尚可,圖像噪聲可接受,可滿(mǎn)足診斷;2分,圖像質(zhì)量較差,噪聲較大,細(xì)節(jié)顯示不清,無(wú)法診斷;1分,圖像質(zhì)量非常差,噪聲大,細(xì)節(jié)顯示不清,無(wú)法診斷。
1.4.2客觀(guān)評(píng)價(jià)主要指標(biāo)包括圖像信號(hào)、噪聲、SNR及CNR。采用圓形或橢圓形ROI,大小約 0.5cm2 ,以隆突水平右側(cè)豎直肌CT值作為圖像信號(hào),標(biāo)準(zhǔn)差作為圖像噪聲,同層面背部皮下純凈脂肪組織CT值作為背景信號(hào),測(cè)量3次取平均值。SNR
圖像信號(hào)/圖像噪聲; CNR= (圖像信號(hào)-背景信號(hào))/圖像噪聲。
1.5統(tǒng)計(jì)學(xué)分析
采用SPSS19.0軟件分析數(shù)據(jù)。定量資料以
表示,分類(lèi)變量以頻數(shù)表示。基本資料、輻射劑量及圖像質(zhì)量客觀(guān)評(píng)分采用獨(dú)立樣本t檢驗(yàn),圖像質(zhì)量主觀(guān)評(píng)分采用Wilcoxon秩和檢驗(yàn)。2位醫(yī)師評(píng)分的一致性行Kappa檢驗(yàn): K?0.75 為一致性好, 0.4
2結(jié)果
2.12組基本資料比較(表1)
LPM組CT平均掃描時(shí)間長(zhǎng)于HPM組,差異有統(tǒng)計(jì)學(xué)意義( Plt;0.05 0;2組年齡、體質(zhì)量及RR差異均無(wú)統(tǒng)計(jì)學(xué)意義(均 Pgt;0.05 )。
2.22組輻射劑量比較(表2;圖1,2)
HPM組CTDI、DLP、ED及過(guò)度掃描距離均顯著高于LPM組,差異均有統(tǒng)計(jì)學(xué)意義(均 Plt;0.05 )。
2.32組圖像質(zhì)量主觀(guān)評(píng)分比較(表3;圖1,2)
2位醫(yī)師對(duì)圖像質(zhì)量主觀(guān)評(píng)分一致性好( K= 0.85)。HPM組整體圖像質(zhì)量?jī)?yōu)于LPM組(運(yùn)動(dòng)偽影評(píng)分:1.6分vs.2.3分,圖像質(zhì)量評(píng)分:3.6分vs.3.1分),差異均有統(tǒng)計(jì)學(xué)意義 (P=0.002,0.004) 。HPM組高RR患者運(yùn)動(dòng)偽影明顯少于LPM組,圖像質(zhì)量評(píng)分明顯優(yōu)于LPM組,差異均有統(tǒng)計(jì)學(xué)意義(均 Plt;0.05, ;而2組低RR患者運(yùn)動(dòng)偽影、圖像質(zhì)量評(píng)分差異均無(wú)統(tǒng)計(jì)學(xué)意義(均 Pgt;0.05 )。
2.42組圖像質(zhì)量客觀(guān)評(píng)價(jià)比較(表4)
2組圖像信號(hào)、噪聲、SNR、CNR差異均無(wú)統(tǒng)計(jì)學(xué)意義(均 Pgt;0.05 )。
表12組基本資料比較

注:HPM為大螺距模式,LPM為小螺距模式,RR為呼吸頻率。
表22組輻射劑量比較
0

注:HPM為大螺距模式,LPM為小螺距模式,CTDI為CT劑量指數(shù),DLP為劑量長(zhǎng)度乘積,ED為有效輻射劑量。
表32組圖像質(zhì)量主觀(guān)評(píng)分比較(分
)

注:HPM為大螺距模式,RR為呼吸頻率,LPM為小螺距模式。
表42組圖像質(zhì)量客觀(guān)評(píng)價(jià)比較 

注:HPM為大螺距模式,LPM為小螺距模式。
3討論
胸部CT檢查是臨床診斷嬰幼兒肺部疾病的首選檢查方法,但對(duì)患兒有一定的輻射傷害。降低管電壓/電流、調(diào)制管電流/電壓及高級(jí)迭代重建算法[5-9]可降低CT輻射劑量。嬰幼兒在行雙源CT胸部檢查時(shí)常采用FlashSpiralHPM及器官-管電流調(diào)節(jié)技術(shù)LPM,這2種掃描模式輻射劑量均在安全范圍之內(nèi)。器官-管電流調(diào)節(jié)技術(shù)又稱(chēng)部分角度掃描模式,在胸部掃描過(guò)程中,當(dāng)X線(xiàn)球管旋轉(zhuǎn)至人體腹側(cè)1/3時(shí),管電流減低,旋轉(zhuǎn)至人體兩側(cè)及背側(cè)2/3時(shí),管電流增加,盡量減低受檢者面?zhèn)人邮茌椛鋭┝浚朔N模式較Flash Spiral HPM的檢查時(shí)間更長(zhǎng)。本研究LPM組平均掃描時(shí)間為 (2.62±1.2)s ,對(duì)自由呼吸狀態(tài)下的嬰幼兒,圖像運(yùn)動(dòng)偽影較明顯,圖像質(zhì)量評(píng)分較低。有研究報(bào)道,HPM雙源CT可減少運(yùn)動(dòng)偽影,提高圖像質(zhì)量[10-12]。本研究也表明,與LPM組相比,HPM組由于檢查時(shí)間較短,在自由呼吸狀態(tài)下嬰幼兒胸部CT圖像質(zhì)量更好,且CT掃描有效劑量較高。原因可能是: ①HPM 的自適應(yīng)動(dòng)態(tài)準(zhǔn)直可降低輻射劑量,但在減少由Z軸過(guò)度掃描效應(yīng)引起的額外輻射照射方面并不理想,本研究中HPM組比LPM組過(guò)度掃描距離多約 60%[(6.54±5.94)cm vs.( 4.02± 2.51)cm] 。 ② 在一定掃描范圍內(nèi)降低管電流,也是LPM輻射劑量更低的重要原因[13]
本研究還發(fā)現(xiàn)自由呼吸狀態(tài)下嬰幼兒的RR對(duì)CT掃描圖像質(zhì)量也非常重要,與 Kim 等[3]研究結(jié)果基本一致。高RR患兒HPM、LPM之間胸部運(yùn)動(dòng)偽影差異更顯著,尤其在膈肌位置,而低RR患兒差異不明顯。HPM組的平均掃描時(shí)間為 (0.37±0.20)s ,而LPM組為 (2.62±1.2)s ;掃描時(shí)間差異造成的膈肌和胸壁運(yùn)動(dòng)偽影是導(dǎo)致這2種模式圖像質(zhì)量差異的主要原因[14-15]。低RR患兒RR最大為25次 /min ,即最短呼吸周期為2.4s,LPH掃描能在嬰幼兒1或2個(gè)呼吸周期內(nèi)完成檢查,這也是低RR患兒HPM、LPMCT掃描圖像質(zhì)量差異不大的重要原因。
本研究存在的局限性: ① 低RR患兒數(shù)量相對(duì)較少,需擴(kuò)大樣本量進(jìn)一步研究; ②2 位醫(yī)師分析圖像時(shí)雖一致性較高,但存在主觀(guān)性; ③ RR測(cè)量和實(shí)際CT圖像采集的時(shí)間間隔較短( lt;1min ),在CT檢查期間RR可能發(fā)生變化; ④ 有效劑量確定困難,且未考慮患兒實(shí)際大小; ⑤ 特定的CT掃描儀和掃描模式進(jìn)行探討,可能需其他CT掃描儀的數(shù)據(jù)來(lái)比較圖像質(zhì)量和輻射劑量。
總之,對(duì)于自由呼吸狀態(tài)下的嬰幼兒,尤其是在CT檢查期間RR較高的患兒,HPM的圖像質(zhì)量常較LPH更好,目前自適應(yīng)動(dòng)態(tài)準(zhǔn)直技術(shù)HPM是最佳選擇。而對(duì)低RR患兒使用LPH行胸部CT檢查,可在減少輻射劑量的同時(shí)保持良好的圖像質(zhì)量。
圖1,2分別為自由呼吸狀態(tài)下大螺距模式( [HPM] 及小螺距模式(LPM)胸部CT圖像注:圖1患兒,男,22個(gè)月,呼吸頻率(RR)27次 /min ,掃描時(shí)間0.28s,無(wú)明顯運(yùn)動(dòng)偽影,圖像質(zhì)量評(píng)分4分,CT容積劑量指數(shù)(CTDIvol) 4.02mGy ,劑量長(zhǎng)度乘積(DLP)72.04mGy?cm (圖1a為肺窗軸位,圖1b為縱隔窗軸位,圖1c為肺窗冠狀位,層厚 2mm ,層距 2mm )。圖2患兒,女,18個(gè)月,RR19次 /min ,掃描時(shí)間 2.9s ,膈肌水平運(yùn)動(dòng)偽影略明顯,圖像質(zhì)量評(píng)分4分,CTDIvol 2.54mGy ,DLP 45.87mGy?cm (圖2a為肺窗軸位,圖2b為縱隔窗軸位,圖2c為肺窗冠狀位,層厚 2mm ,層距 2mm )

[參考文獻(xiàn)]
[1]LELL M M,MAYM,DEAK P,etal.Radkow,high-pitch spiralcomputed tomography:effect onimage quality and radiation dose in pediatric chest computed tomography,invest[J]. Radiol,2011,46(2):116-123.
[2]KETELSEN D,BUCHGEISTE R M,F(xiàn)ENCHEL M,et al. Automated computed tomography dosesaving algorithm
toprotect radiosensitive tissues:estimation of radiation exposure and image quality considerations[J]. Inv Rad, 2012,47(2):148-152.
[3]KIM S H,CHOI Y H,CHO H H,et al.Comparison of image quality and radiation dose between high-pitch modeand low-pitch mode spiral chest CT in small uncooperative children:the effect of respiratory rate[J].
EurRadiol,2016,26(4):1149-1158.
[4]DEAK P D,SMAL Y,KALENDER W A.Multisection CT protocols:sex-and age-specific conversion factors used to determine effective dose from dose-length product[J]. Radiology,2010,257(1):158-166.
[5]唐雷,劉莉,曾憲春,等.能譜純化結(jié)合高級(jí)模擬迭代重 建在成人胸部低劑量CT中的應(yīng)用[J].中國(guó)介入影像與 治療學(xué),2018,15(8):496-500.
[6]LEE SW,KIM Y,SHIM S S,et al. Chang,image quality assessment of ultra low-dose chest CT using sinogramaffirmed iterative reconstruction[J].Eur Radiol,2014, 24(4):817-826.
[7]王苑丁,康德強(qiáng),白玫,等.不同強(qiáng)度高級(jí)迭代重建算法 在不同噪聲水平下對(duì)圖像質(zhì)量的影響:體模研究[J].中 國(guó)醫(yī)學(xué)影像技術(shù),2019,35(6):914-919.
[8]GORDIC S,DESBIOLLES L,SEDLMAIR M,et al.Optimizing radiation dose byusing advanced modelled iterative reconstruction in high-pitch coronary CT angiogra phy[J]. Eur Radiol,2016,26:459-468.
[9]HWANG H J,SEO JB,LEE H J,et al. Low-dose chest computedtomography with sinogram-affirmed iterative reconstruction,iterative reconstruction in image space, and filtered back projection:studies on image quality [J].JComput Assist Tomogr,2013,37(4):610-617.
[10]閆景偉.雙源CT大螺距掃描模式下 70kV 管電壓結(jié)合 迭代算法用于CT造影檢查對(duì)圖像質(zhì)量及輻射劑量的影 響[J].實(shí)用醫(yī)學(xué)影像雜志,2021,22(6):576-579.
[11]TABARI A,PATINO M,WESTRA S J,et al. Initial clinical experience with high-pitch dual-source CT as arapid technique for thoracoabdominal evaluation in awake infants and young children[J].Clin Radiol, 2019,74(12):977.e9-977.e15.
[12]李騁,孫英麗,金係,等.大螺距掃描模式在COPD患者 肺CT低劑量篩查中的初步應(yīng)用[J].臨床放射學(xué)雜志, 2018,37(8):1381-1385.
[13]王光大,時(shí)高峰,楊麗,等.X-care技術(shù)在胸部CT掃描 中對(duì)患者敏感器官保護(hù)作用的研究[J].中國(guó)醫(yī)療設(shè) 備,2017,32(1):84-86.
[14]SCHULZ B,JACOBI V,BEERES M,et al. Quantitative analysis of motion artifacts in high-pitch dual-source computed tomography of the thorax[J].J Thorac Imaging,2012,27(6):382-386.
[15]BODELLEB,F(xiàn)ISCHBACH C,BOOZ C,et al.Freebreathing high pitch 80kVp dual-source computed tomographyof the pediatric chest:imagequality,presence ofmotion artifacts and radiation dose[J].Eur JRadiol,2017,89:208-214.
(收稿日期 2023-06-27)
中國(guó)中西醫(yī)結(jié)合影像學(xué)雜志2025年5期