Comparison of measurements of anterior chamber angle via anterior segment optical coherence tomography and ultrasound biomicroscopy

2020-09-10 02:57:24ZiYanYuTingHuangLuLuBoQu
World Journal of Clinical Cases 2020年15期

Zi-Yan Yu, Ting Huang, Lu Lu, Bo Qu

Zi-Yan Yu, Lu Lu, Bo Qu, Department of Ophthalmology, Fourth Affiliated Hospital of China Medical University, Eye Hospital of China Medical University, Key Laboratory of Lens Research of Liaoning Province, Shenyang 110005, Liaoning Province, China

Ting Huang, Department of Cataract, Aier Mega Ophthalmic Hospital of Chongqing,Chongqing 400060, China

Abstract

Key words:Anterior segment-optical coherence tomography;Primary angle closure glaucoma;Anterior chamber angle;Angel opening distance at 500 m;Trabeculo-iris space area at 500 m2

INTRODUCTION

Glaucoma is an irreversible optic neuropathy with the loss of visual field and decrease of vision.There are no obvious clinical manifestations in the early stage of glaucoma;typical visual field defects or progressive loss of visual acuity and other symptoms may appear in the moderate or severe phase of glaucoma[1,2].Approximately 80% of aqueous humor flows out through the anterior chamber angle, when aqueous humor outflow obstruction is caused by abnormality of anterior chamber angle structure or function, and compensatory imbalance of aqueous humor leads to increased intraocular pressure.

At present, gonioscopy, anterior segment optical coherence tomography (AS-OCT),and ultrasound biomicroscopy (UBM) are commonly used to evaluate anterior chamber angle[3-5].AS-OCT and other non-contact imaging modalities provide a quick,user-friendly, and objective non-invasive method of assessing the anterior segment parameters and anterior chamber angle, are well tolerated by the patient, and correlate well with the information provided by gonioscopy[6,7].

In this study, we evaluated the agreement, difference, and correlation of chamber angle parameters such as angel opening distance at 500 μm (AOD500) and trabeculoiris space area at 500 μm2(TISA500) measured by AS-OCT and UBM.

MATERIALS AND METHODS

Study subjects

Forty-five subjects (15 males and 30 females;age range, 50–80 years) without glaucoma who visited the Fourth Affiliated Hospital of China Medical University from October 2015 to August 2016 were selected.All subjects underwent routine ophthalmological examinations (including visual acuity, intraocular pressure, integrated optometry, slit lamp examination, and fundus examination) and AS-OCT and UBM.None of the subjects had an intraocular surgery or laser procedure, and none of the subjects had evidence of glaucoma, dry eye disease, ocular inflammatory activity, or other ocular diseases.

This study was approved by the Fourth Affiliated Hospital of China Medical University Committee on Human Research.The registry’s URL is www.medresman.org and the registration number is ChiCTR1900021378.All study procedures adhered to the Declaration of Helsinki for research involving human participants.All participants provided informed consent after explanation of possible consequences of the study to them.

AOD500 and TISA500 evaluation procedures

The patients were examined in the order of AS-OCT (sitting position) and UBM(supine position).The interval of examinations was about 5 min;all the examinations with the same instrument were performed three times in the same brightness environment and by the same operator.The average values of AOD500 and TISA500 of four quadrants for each eye were recorded.

Statistical analysis

All descriptive analyses, difference analyses, and correlation analyses were performed with SPSS version 19.0, and agreement analysis was performed with MedCalc version 15.0.All descriptive results are expressed as the mean ± SD.The differences in the two instruments were analyzed by pairedt-test if the data were in accordance with the normal distribution and the variance was homogeneous;otherwise, nonparametric tests were used.Correlations and agreements between the instruments were performed by Pearson correlation analysis and Bland–Altman analysis.P<0.05 was considered statistically significant.

RESULTS

A total of 15 male subjects (24 eyes) and 30 female subjects (48 eyes) were examined by UBM and AS-OCT for anterior chamber angles parameters (AOD500 and TISA500),with a mean age of 59.47 ± 14.481 years.

AOD500 and TISA500 measured by AS-OCT and UBM

AOD500 at 90° (12 o’clock position) was 0.2453 ± 0.2566 mm by AS-OCT and 0.1642 ±0.1502 mm by UBM.AOD500 at 0° (3 o’clock position) was 0.2493 ± 0.2065 mm by ASOCT and 0.2090 ± 0.1344 mm by UBM.AOD500 at 270° (6 o’clock position) was 0.2828± 0.2587 mm by AS-OCT and 0.1914 ± 0.1410 mm by UBM.AOD500 at 180° (9 o’clock position) was 0.2661 ± 0.2300 mm by AS-OCT and 0.2194 ± 0.1523 mm by UBM.

TISA500 at 90° (12 o’clock position) was 0.0912 ± 0.0884 mm by AS-OCT and 0.0749± 0.0630 mm by UBM.TISA500 at 0° (3 o’clock position) was 0.0997 ± 0.0716 mm by AS-OCT and 0.0982 ± 0.0676 mm by UBM.TISA500 at 270° (6 o’clock position) was 0.1066 ± 0.0941 mm by AS-OCT and 0.0842 ± 0.0625 mm by UBM.TISA500 at 180° (9 o’clock position) was 0.1071 ± 0.0773 mm by AS-OCT and 0.1004 ± 0.0765 mm by UBM.

For AOD500 and TISA500 of the four quadrants of the chamber angles,measurements by AS-OCT were greater than those by UBM.AOD500 measured by AS-OCT from wide to narrow were 6 o’clock, 9 o’clock, 3 o’clock, and 12 o’clock positions.AOD500 measured by UBM from big to small were 9 o’clock, 3 o’clock, 6 o’clock, and 12 o’clock positions.TISA500 measurements by AS-OCT from wide to narrow were 9 o’clock, 6 o’clock, 3 o’clock, and 12 o’clock positions.TISA500 measurements by UBM from big to small were 9 o’clock, 3 o’clock, 6 o’clock, and 12 o’clock positions.

Differences in AOD500 and TISA500 measured by AS-OCT and UBM

The measurements of AOD500 and TISA500 using the two methods were not normally distributed, and the homogeneity of variance (Mann–Whitney test) and the differences in them were compared using nonparametric tests.There was no significant differences between the two methods in measuring AOD500 of the four quadrants of the chamber angles (P1=0.110,P2=0.633,P3=0.078, andP4=0.474).There was no significant differences between the two methods in measuring TISA500 of the four quadrants of the chamber angles (P1=0.584,P2=0.889,P3=0.297, andP4=0.550).Comparing the AOD500 and TISA500 measured by AS-OCT and UBM, the results were not significantly different (P>0.05).

Correlations of AOD500 and TISA500 measured by the two methods

The AOD500 measurements by AS-OCT and UBM at 90° (r=0.562,P<0.001), 0° (r=0.671,P<0.001), 270° (r=0.635,P<0.001), and 180° (r=0.720,P<0.001) were highly correlated (Figure 1).

The TISA500 measurements by AS-OCT and UBM at 90° (r=0.466,P<0.001), 0° (r=0.652,P<0.001), 270° (r=0.507,P<0.001), and 180° (r=0.545,P<0.001) were highly correlated (Figure 2).

Agreements of AOD500 and TISA500 measured by the two methods

Agreement of the AOD500 measurements by the two methods was analyzed by Bland–Altman diagram at the four quadrants of the anterior chamber angles(Figure 3).At 90°, four (2.78%) of 144 points were outside the confidence interval.At 0°, four of 144 points were outside the confidence interval.At 270°, three (2.08%) of 144 points were outside the confidence interval.At 180 °, four of 144 points were outside the confidence interval.The Bland–Altman diagram of AOD500 measurements at any quadrants of the chamber angles showed that >95% of the data points were in the 95%agreement interval and the confidence interval was relatively narrow, with no significant difference, and could be clinically accepted.Bland–Altman diagram analysis showed that the results of AOD500 measurements at four quadrants of the chamber angles measured by the two methods were in good agreement.

Agreements of the TISA500 measurements by the two methods were analyzed by Bland–Altman diagram at the four quadrants of the anterior chamber angles(Figure 4).At 90°, four (2.78%) of 144 points were outside the confidence interval.At 0°, five (3.47%) of 144 points were outside the confidence interval.At 270°, three(2.08%) of 144 points were outside the confidence interval.At 180°, five (3.47%) of 144 points were outside the confidence interval.The Bland–Altman diagram of TISA500 measurements at any quadrants of the chamber angles showed that >95% of the data points were in the 95% agreement interval and the confidence interval was relatively narrow, with no significant difference, and could be clinically accepted.Bland–Altman diagram analysis showed that the results of TISA500 measurements at four quadrants of the chamber angles measured by the two methods were in good agreement.

DISCUSSION

Microstructures of the anterior chamber angle (such as iris root, scleral protrusion, and trabecular meshwork) can be observed to estimate the degree of opening of the anterior chamber angle, and its structure and function can be estimated by gonioscopy[8,9].Observation of a scleral spur suggests complete exposure of the functional trabecular meshwork, and this is a marker in imaging examination and the first step in observation[10].AOD500 and TISA500 measurements are affected by body position, eye position, whether there is contact or pressure on the eye during measurement, and whether front and rear boundary measurements are accurate[11,12].Under the influence of gravity, the cornea can be compressed by the upper eyelid, and the normal anterior chamber angle shows that the top position is narrowest and the bottom position is widest, followed by nasal and temporal position[13].

At present, objective methods of visualizing anterior chamber angle are UBM and AS-OCT.AOD500 and TISA500 measurements by AS-OCT were, respectively, 0.2453 ±0.2566 mm and 0.0912 ± 0.0884 mm at 90°, 0.2828 ± 0.2587 mm and 0.1066 ± 0.0941 mm at 270°, 0.2493 ± 0.2065 mm and 0.2661 ± 0.2300 mm at 0°, and 0.0997 ± 0.0716 mm and 0.1071 ± 0.0773 mm at 180°.AS-OCT is a non-contact examination with the subject in the sitting position, and it is not affected by body and eye positions[14].In AOD500 and TISA500 measurements, the top position angle is narrowest and coincides with the physiological state.In AOD500 measurements, the bottom position angle is widest and coincides with the physiological state.In TISA500 measurements, related to the measurement error, the 270° angle is widest but not much different from the bottom position angle.There are many reasons for the measurement error.The eyeball may be completely exposed while measuring, which can exert pressure on the eyeball, thus affecting the accuracy of measurement.The measurements are automatically identified by computer, resulting in inaccuracy of the positioning boundaries.

Figure 1 Correlation diagrams of angel opening distance at 500 μm measured by the two instruments of the four quadrants of the chamber angles.

AOD500 and TISA500 measurements by UBM were, respectively, 0.1642 ± 0.1502 mm and 0.0749 ± 0.0630 mm at 90° position, 0.1914 ± 0.1410 mm and 0.0842 ± 0.0625 mm at 270°, 0.2090 ± 0.1344 mm and 0.2194 ± 0.1523 mm at 0°, and 0.0982 ± 0.0676 mm and 0.1004 ± 0.0765 mm at 180°.In AOD500 and TISA500 measurements, the top position angle is narrowest, the 180° position angle is widest, and the 0° position takes second place.Considering that UBM is performed with the patient in the supine position, change of barycenter leads to redistribution of aqueous humor;the width of the horizontal anterior chamber angle increases, and the width of the lower anterior chamber angle decreases.When measuring in one direction, the patient’s eyeball rotates and eye position changes in that direction, which can cause measurement error.It is a contact examination and has a greater possibility of eye compression, both of which can affect the measurement results.Moreover, UBM resolution is lower, and manual measurement can introduce human error.

Comparing the degree of openness of the anterior chamber angle can determine opening and closing and degree of stenosis of the anterior chamber angle.Moreover, it can be used to compare preoperative and postoperative chamber angle changes[15-17].By comparing the AOD500 and TISA500 measurementsviaUBM and AS-OCT, we conclude that the agreement and correlation are excellent, and there is no significant difference between them, proving that they are interchangeable.Previous studies also support this viewpoint[1,18-21].Compared with UBM, AS-OCT has better repeatability,noninvasiveness, rapidity, high resolution, and automatic analysis.These features can avoid measurement errors caused by eye compression, eyeball rotation, and body position change, making anterior chamber angle parameters more coincident with those under physiological conditions.AS-OCT can also avoid human measurement error, making the data in good repeatability.As a non-contact device, with short measure time, fast and clear imaging, and easily follow-up, AS-OCT can be widely used in glaucoma screening, open eye injury, and other diseases.However, UBM can be used to diagnose ciliary body detachment, ciliary body tumor, peripheral retinal detachment, and other diseases, while AS-OCT cannot.

The effect of intragroup differences on the results was considered.Age, sex, and laterality of the eye were not distinguished.The measurements were not classified based on the width of anterior chamber angle and status of diseases.There were only 72 eyes, and the data did not satisfy a normal distribution.The influence of the above factors on the measurement results should be investigated further using a larger study sample size.

In conclusion, there is a good correlation and agreement between AOD500 and TISA500 measurements by AS-OCT and UBM.The two methods are interchangable for measurement of the anterior segment angle and cannot be completely replaced by each other.AS-OCT can be used as the first choice for early glaucoma screening, and can be used for follow-up of glaucoma surgery.

Figure 2 Correlation diagrams of trabeculo-iris space area at 500 μm2 measured by the two instruments of the four quadrants of the chamber angles.

Figure 4 Bland-Altman charts of trabeculo-iris space area at 500 μm2 at four quadrants of the chamber angles measured by the two devices.

ARTICLE HIGHLIGHTS

Research background

Glaucoma is an irreversible optic neuropathy with the loss of visual field and decrease of vision.A quick, user-friendly, and non-invasive method of assessing the anterior segment parameters is essential for early screening and diagnosis of glaucoma.

Research motivation

The main topic is to compare the difference between anterior chamber angle parameters measured by different methods.

Research objectives

To investigate the agreement, difference, and correlation of chamber angle parameters such as angel opening distance at 500 μm (AOD500) and trabeculo-iris space area at 500 μm2(TISA500) measured by anterior segment optical coherence tomography (ASOCT) and ultrasound biomicroscopy (UBM) in normal subjects.

Research methods

Both AS-OCT and UBM were performed to measure AOD500 and TISA500 in 45 subjects (72 eyes).All subjects without glaucoma were collected from October 2015 to August 2016 at the Ophthalmology Department of the Fourth Affiliated Hospital of China Medical University.Data of the two groups (AOD500 and TISA500) were compared by nonparametric tests.Pearson correlative analysis and Bland–Altman analysis were used to compare the correlation and agreement.

Research results

There were no significant differences between AS-OCT and UBM in measuring AOD500 and TISA500 of the four quadrants of the anterior chamber angle.There was a high correlation in measuring AOD500 and TISA500.There was a good agreement in measuring AOD500 and TISA500.

Research conclusions

There is a high correlation and agreement between AOD500 and TISA500 measurements by AS-OCT and UBM.They are interchangeable under some circumstances.AS-OCT proves to be a better early screening tool for glaucoma.

Research perspectives

AS-OCT and UBM are interchangable for AOD500 and TISA500 measurements and cannot be completely replaced by each other.AS-OCT can be used as the first choice for early glaucoma screening, especially for primary open-angle glaucoma patients.

主站蜘蛛池模板: 这里只有精品在线播放| 国产一区二区丝袜高跟鞋| 中文字幕久久波多野结衣| 中文字幕首页系列人妻| 国产美女免费网站| 99偷拍视频精品一区二区| 97免费在线观看视频| 久草视频一区| 久久久亚洲色| 日本午夜影院| 天天综合网亚洲网站| 国产日韩av在线播放| 国产麻豆91网在线看| 欧美色视频在线| 欧美a在线看| 亚洲精品欧美日本中文字幕| 国产人成网线在线播放va| 真实国产精品vr专区| 亚洲开心婷婷中文字幕| 不卡无码网| 十八禁美女裸体网站| 99草精品视频| 久久国产精品电影| 国产日韩精品欧美一区灰| 亚洲一区毛片| 亚洲精品卡2卡3卡4卡5卡区| 亚洲高清资源| 91在线精品麻豆欧美在线| 久久中文无码精品| 国产精品主播| 日韩无码黄色| 92午夜福利影院一区二区三区| 国产午夜精品鲁丝片| 亚洲视频二| 高清色本在线www| 激情乱人伦| 99久久亚洲精品影院| 国产浮力第一页永久地址| 在线无码私拍| 亚洲有无码中文网| 日本AⅤ精品一区二区三区日| 国产一二三区视频| 久久综合亚洲鲁鲁九月天| 澳门av无码| 国产精品成人免费视频99| 亚洲日韩国产精品综合在线观看| 综合久久久久久久综合网| 国产无码高清视频不卡| 国产成人精品免费视频大全五级| 亚洲欧洲日产国产无码AV| 午夜成人在线视频| 亚洲国产成人久久77| 精品中文字幕一区在线| 97在线碰| 人人妻人人澡人人爽欧美一区| 国产流白浆视频| 久精品色妇丰满人妻| 深爱婷婷激情网| 国产96在线 | 亚洲精品高清视频| 美女内射视频WWW网站午夜| 免费视频在线2021入口| 九九香蕉视频| 国产乱人激情H在线观看| 亚洲人成色在线观看| 国产99欧美精品久久精品久久| 国产欧美中文字幕| 亚洲国产亚综合在线区| 亚洲综合一区国产精品| 国产男人天堂| 免费国产高清精品一区在线| 欧美伦理一区| 国产日韩丝袜一二三区| 色婷婷亚洲综合五月| 亚洲a级在线观看| 国产成人精品男人的天堂| 青青青国产在线播放| 免费毛片视频| 欧美日韩精品一区二区视频| 91精品免费高清在线| 经典三级久久| 亚洲一区二区三区麻豆|