Comparison of imo and Humphrey field analyzer perimeters in glaucomatous eyes

2021-12-17 02:42:30YoshinoriNakaiKyokoBesshoYukoShonoKaoriTaokaYoshihideNakai
International Journal of Ophthalmology 2021年12期

Yoshinori Nakai, Kyoko Bessho, Yuko Shono, Kaori Taoka, Yoshihide Nakai

Tokai Eye Clinic, 399-Hadokoro-cho, Tsu 514-0009, Japan

Abstract

● KEYWORDS: imo perimeter; visual field diagnosis;diagnostic validation

INTRODUCTION

An automated visual field analyzer (perimeter) is indispensable for glaucoma diagnosis and follow-up.However, the conventional equipment is heavy, and a large amount of installation space is needed, and the examining room must be dark. For an examination, patients must sit in front of the visual field analyzer; therefore, bedside examinations are not possible. Recently, however, a head-mounted device, the imo perimeter (Crewt Medical Systems, Inc., Tokyo, Japan),has been developed to be lightweight (1.8 kg) and to be used in a brightly lit room. It can help clinicians perform visual field inspections and observe the central and peripheral visual fields of up to 30 degrees in both eyes simultaneously within a short time period (Figure 1)[1-11].

We compared glaucoma detection sensitivities, diagnostic abilities in addition to the correlation of each parameter and examination time of the Humphrey field analyzer (HFA;Humphrey Instruments, Dublin, CA, USA), which is currently in wide use, with those of imo perimeters.

With the imo perimeter, the left and right pupils are monitored individually with a near-infrared camera for fixation disparity,which reduces the examination time (Figure 2). In addition,with the imo perimeter, a transmissive liquid crystal display at full high-definition resolution and a high-brightness lightemitting diode backlight are used in the same conditions as those of the HFA perimeter. Furthermore, targets are shown to the left and right eyes of subjects separately but simultaneously by independent left and right optical systems. In the visual field test, the fixed target is visually fused, and patients fixate on the target with both eyes open (Figure 3).

The imo perimeter has a mode of 24 plus (78 test points) to which a 10-2 visual field is added (so that it is partially like the HFA perimeter with its visual fields of 30-2, 24-2, and 10-2),as well as a mode of 24 plus 1 (36 points), in which the points are preferentially focused on sites in which disease is likely to occur. To examine the correlation of the measurements made with the HFA and imo perimeters for mild to severe glaucoma, we used a 30-2 examination program (widely used in glaucoma diagnosis), in which a 30-degree visual field was measured with intervals of 6 degrees.

SUBJECTS AND METHODS

Figure 1 Head-mounted imo perimeter weighing 1.8 kg It can be used in a well-lit room, and it enables both eyes to be examined quickly and simultaneously.

Figure 2 Simultaneous monitoring of binocular fixation by imo perimeter, with automatic tracking correction (from side to side and up and down), and pupil diameter measurement.

Figure 3 The test target was presented randomly to either eye under a nonocclusion condition, and the patient was not aware of which eye was being tested.

Ethical Approval This study was conducted in compliance with the tenets of the Helsinki Declaration, and all participants provided written informed consent.Visual field tests were performed with both the HFA and imo perimeters on the same day for each patient on 128 eyes of 64 patients with glaucoma and 40 normal eyes of 20 healthy people, all of whom were examined at the Tokai Eye Clinic,Tsu, Japan, between October 2018 and February 2019. Patients with glaucoma who completely understood the contents of the visual field tests and demonstrated good visual fixation in testing were selected.

Table 1 Characteristics of the participants in the study

Using Pearson’s correlation coefficient, we compared the parameters—the mean deviation and defect (MD), the pattern standard deviation (PSD), and the visual field index (VFI)—as measured by both the HFA and imo perimeters. We calculated the detection sensitivity by using Bland-Altman analysis, the diagnostic ability by the area under the receiver operating characteristic curve (AUC), and, with thettest, the examination time.

Background of Patients with Glaucoma The participants with glaucoma had a spherical equivalent of -6 to +3 D,astigmatism of ±2.5 D, and best-corrected visual acuity(logMAR) of 0.2 to 0.08. They had not undergone eye surgery in the past and had no eye disease except glaucoma. The patients were 18 to 72 years of age, with an average age of 59.5±13.5y; 25 were men, and 39 were women (Table 1).

RESULTS

In the 40 normal eyes, the mean examination times were 12±1.4min with the HFA perimeter and 8.3±1.8min with the imo perimeter (P<0.001). In the 128 glaucomatous eyes, the mean examination times were 16.2±2.5min with the HFA perimeter and 12.0±2.5min with the imo perimeter (P<0.001).Thus the examinations of both the normal eyes and the eyes with glaucoma were performed in a significantly short time(P<0.001; Table 2).

Pearson’s correlation coefficients with the HFA and imo perimeters in the 128 eyes with glaucoma were 0.886(P<0.001) for MD,0.814(P<0.001) for PSD, and 0.871 (P<0.001) for VFI; strong correlations were thus noted (Figure 4). In the 40 normal eyes,the correlation coefficients were 0.248(P<0.124) for MD,0.111(P<0.496) for PSD, and 0.028(P<0.864) for VFI.

Figure 4 In 128 glaucomatous eyes, the imo perimeter and the HFA perimeter’s measurements of MD (r=0.886, P<0.001), PSD (r=0.814,P<0.001), and VFI (r=0.871, P<0.001) were strongly and positively correlated.

Figure 5 Displays of the visual field by the HFA and imo perimeters In case 1 (mild glaucoma), there was no difference in the displays; in case 2, 3 and 4 (advanced glaucoma), however, the differences were obvious.

Table 2 Examination time of imo and HFA perimeters min, mean±SD

To compare the MD, PSD, and VFI measured by both the perimeters, we performed the Bland-Altman analysis because even if those parameters of the HFA and imo perimeters were correlated, we suspected a difference would reflect a worse rate of glaucoma. Figure 5 illustrates the results of a comparison of the visual field findings of the HFA and imo perimeters in patients 1, 2, 3, and 4. Patient 1, in whom glaucoma was mild,showed no difference in parameters, whereas patients 2, 3 and 4, in whom the glaucoma was advanced, did show differences.The results of the Bland-Altman analysis are shown in Figure 6, in which the vertical axis reflects the difference between the imo and HFA perimeters and the horizontal axis shows the average values of the HFA and imo perimeters. The distribution was fan-shaped, whereby the worse the glaucoma was (horizontal axis), the greater was the difference in results between the imo and HFA perimeters. This finding reflects a proportional error.

The diagnostic ability of both perimeters was examined with AUC (Figure 7). For MD, the AUCs were 0.911 for the imo perimeter and 0.819 for the HFA perimeter; for PSD, they were 0.885 for the imo perimeter and 0.890 for the HFA perimeter;and for VFI, they were 0.882 for the imo perimeter and 0.872 for the HFA perimeter. Both the imo and HFA perimeters demonstrated high diagnostic ability, and no significant difference in diagnostic ability was noted between them.

DISCUSSION

In addition to 30-2, 24-2, and 10-2 perimetry inspection modes, which the HFA perimeter has, the imo perimeter has a mode termed “24 plus” (78 test points). This mode consists of some of the test points of 10-2 added to those of 24-2 (36 test points), in which test points are preferentially arranged at sites of the 24 plus mode in which disease is likely to occur. An algorithm termed “AIZE-Rapid” (Crewt Medical Systems) can be used to further shorten inspection time. The imo perimeter is highly reliable because it has two independent optical systems that can both perform a visual field examination with both the patient’s eyes open and perform eye tracking of the pupil.

Figure 6 Bland-Altman analysis was used to compare MD, PSD, and VFI measurements by the HFA perimeter and the imo perimeter For glaucoma eye a proportional error was observed. The results show the fan-shaped distribution, where the worse is the glaucoma (horizontal axis), the greater is the difference between imo and HFA results.

Figure 7 The diagnostic abilities of both the HFA and the imo perimeters were significant; AUC was 0.8 or greater in the MD, PSD, and VFI measurements, and the maximum value of the AUC was 0.911 for the MD measurement by the imo perimeter.

In this study, 128 eyes of 64 glaucoma patients and 40 eyes of 20 healthy people were examined with the 30-2 mode,which is widely currently used in glaucoma testing. The current basic measurement of static visual field consists of examining approximately 70 measurements points within the central 30 degrees with intervals of 6 degrees, as indicated by the HFA perimeter 30-2 mode. The arrangement of the gridlike measurement points with intervals of 6 degrees has poor sensitivity for early glaucoma detection and macular disease;therefore, the use of an additional macular mode such as 10-2 is recommended.

The purpose of this study was to investigate the correlation between imo and HFA perimeter and the reliability of imo perimeter. Therefore, in many cases, abnormalities appeared early within 30 degrees of the visual field center. In 30-2,comparison of measurement time of two perimeters, the correlation of each parameter, Bland-Altman analysis, and diagnostic ability with AUC were examined.

In the comparison between the widely used HFA and the Octopus perimeters[12]and in comparison with the KOWA AP-7000 perimeter[13]reported strong correlations. With regard to the effect of different arrangements, these investigators reported that there was no substantial difference between the 30-2 and 24-2 modes to the HFA perimeter[14-15]. As our objective was to examine the correlation between the imo and HFA perimeters and the reliability of the imo perimeter, we compared the measurement times of both perimeters by using the 30-2 mode. This is the mode in which abnormalities caused by many diseases are expressed earliest within the central 30 degrees of the visual field. We checked the correlation of each parameter, performed a Bland-Altman analysis, and used the AUC to examine the glaucoma diagnostic ability of both perimeters.

Khouryet al[14]compared the examination times of the HFA with 24-2 and 30-2 modes, demonstrating an average of 10min and 24s with the 24-2 mode and an average of 14min and 24s with the 30-2 mode. Our examinations of 128 glaucomatous eyes and 40 normal eyes were significantly shorter with the imo perimeter than with the HFA perimeter:shorter by 4.2min for glaucomatous eyes and by 3.8min for normal eyes (Table 2). Because the imo perimeter is capable of measuring both eyes simultaneously in a brightly lit room,the burden on patients was reduced. It is thought that the examination time with the imo perimeter is shorter because it individually monitors the fixation of the left and right pupils and automatically corrects fixation disparity and because the AIZE-Rapid algorithm further speeds up examination.

Pearson’s correlation coefficients of MD, PSD, and VFI were 0.886, 0.814, and 0.871 (allP<0.001), respectively, and the correlations were significantly positive (Figure 4). In normal eyes, the reason that the MD, PSD, and VFI were 0.248, 0.111,and 0.028 and were lower than those in glaucomatous eyes was probably because the parameter values of normal eyes was dense. Even though there was a positive correlation between the two perimeters, we performed a Bland-Altman analysis to examine whether there was any difference as a result of the progression of glaucoma. The analysis showed that the difference between the imo and HFA perimeters tended to increase as glaucoma progressed, and a proportional error was observed in MD, PSD, and VFI in all cases (Figure 5). Thus caution is required in the determinations of these parameters because the difference between results obtained with the imo and HFA perimeters increases as the visual field disturbance progresses. This demonstrates that the measurements obtained from the imo may indicate a higher rate of worsening of glaucoma than previously realized and that this rate may increase. However, there may be some cases in which the measurements improve.

The diagnostic ability of both perimeters was examined in receiver operating characteristic analysis (Figure 6). The highest AUC (0.911) was observed in MD with the imo perimeter(95% CI: 0.867 to 0.955). All MD, PSD, and VFI values exceeded 0.8 for both the imo and HFA perimeters, and the diagnostic abilities of the two perimeters were all significant (P<0.001).Moreover, the AUC of the MD was significantly higher for the imo perimeter than for the HFA perimeter.

In addition to the head-mounted configuration, other advantages of the imo perimeter are its light weight and small size, which facilitate easy handling. These features not only allow patients to be examined in bed but also expand the range of clinical applications, including the conducting of examinations in a small space[16-19].

The National Aeronautics and Space Administration is studying and elucidating symptoms of space flight-associated neuroocular syndrome (SANS), which occur in the whole body,nerves, and eyes during long space flights. Symptoms such as optic disc edema, globe flattening, choroidal and retinal folds,hyperopic refractive error shifts, and infarcts in nerve fiber layers have been reported to occur during long-term residence in a space station[20-21]. The imo perimeter will be a useful apparatus for performing visual field examinations in a space station.

ACKNOWLEDGEMENTS

Conflicts of Interest:Nakai Y, None; Bessho K, None;Shono Y, None; Taoka K, None; Nakai Y, None.

主站蜘蛛池模板: 九色综合视频网| 日韩天堂网| 噜噜噜久久| 麻豆AV网站免费进入| 亚洲国产天堂久久综合226114| 久久亚洲高清国产| 精品视频在线观看你懂的一区| 亚洲综合天堂网| 久久一级电影| 97se亚洲综合在线| 欧美日韩成人在线观看| 国产福利2021最新在线观看| 久久国产香蕉| a免费毛片在线播放| 久久综合激情网| 一本大道无码高清| 97国产成人无码精品久久久| 2019年国产精品自拍不卡| 国产一级毛片在线| 思思热在线视频精品| 国产亚洲欧美另类一区二区| 免费高清毛片| 伊人久久久久久久| 大香网伊人久久综合网2020| 成人va亚洲va欧美天堂| 成人欧美在线观看| 精品無碼一區在線觀看 | 九九热精品免费视频| 国产乱子伦精品视频| 欧美高清国产| 免费无码AV片在线观看国产| 色婷婷电影网| 国产激爽大片高清在线观看| 国产99热| 伊人网址在线| 欧美h在线观看| 国产鲁鲁视频在线观看| 日韩第九页| 91成人在线观看| 色香蕉影院| 色综合天天视频在线观看| 国产www网站| 国产欧美日韩一区二区视频在线| 性色一区| 精品国产成人国产在线| 欧美日韩亚洲国产| 一本久道久综合久久鬼色| 一级毛片在线播放| 欧美一级夜夜爽| 免费AV在线播放观看18禁强制| 婷婷色丁香综合激情| 无码国产偷倩在线播放老年人| 午夜精品福利影院| 九九久久99精品| 国产丝袜91| 国内精品视频| 国产女同自拍视频| 乱系列中文字幕在线视频| 五月婷婷丁香综合| 欧美亚洲日韩中文| 国产乱人伦AV在线A| 爽爽影院十八禁在线观看| 亚洲制服丝袜第一页| 国产成人在线小视频| 国产精品吹潮在线观看中文| 91精品久久久无码中文字幕vr| 美女内射视频WWW网站午夜 | 国产成人免费| 欧美日韩第二页| 久久这里只有精品66| 国产美女自慰在线观看| 国产高清色视频免费看的网址| 国产经典在线观看一区| 中文纯内无码H| 国产日韩久久久久无码精品| 亚洲欧洲综合| 亚洲欧美日韩中文字幕在线一区| 久久精品人人做人人综合试看| 国产精品私拍在线爆乳| 天堂av高清一区二区三区| 91视频精品| 国产午夜无码片在线观看网站|