How to secure the connection between thoracostomy tube and drainage system?

2014-03-20 07:39:07KaKiPatLiKitShingJohnWongYauHangHenryWongKaLokChengFungLingSoChuLeungLauChakWahKam
World journal of emergency medicine 2014年4期

Ka Ki Pat Li, Kit Shing John Wong, Yau Hang Henry Wong, Ka Lok Cheng, Fung Ling So, Chu Leung Lau, Chak Wah Kam

1Department of Accident and Emergency Medicine, Pok Oi Hospital, Au Tau, Yuen Long, Hong Kong, China

2Department of Accident and Emergency Medicine, Tuen Mun Hospital, Tuen Mun, Hong Kong, China

3Clinical Skills Training Centre, Tuen Mun Hospital, Tuen Mun, Hong Kong, China

Corresponding Author:Ka Ki Pat Li, Email: lkk282@ha.org.hk

How to secure the connection between thoracostomy tube and drainage system?

Ka Ki Pat Li1, Kit Shing John Wong1, Yau Hang Henry Wong2, Ka Lok Cheng3, Fung Ling So1, Chu Leung Lau1, Chak Wah Kam1

1Department of Accident and Emergency Medicine, Pok Oi Hospital, Au Tau, Yuen Long, Hong Kong, China

2Department of Accident and Emergency Medicine, Tuen Mun Hospital, Tuen Mun, Hong Kong, China

3Clinical Skills Training Centre, Tuen Mun Hospital, Tuen Mun, Hong Kong, China

Corresponding Author:Ka Ki Pat Li, Email: lkk282@ha.org.hk

BACKGROUND:Thoracostomy tube insertion is one of the common bedside procedures in emergency medicine and many acute specialties. Dislodgement of thoracostomy tube from the connection tube of chest drainage system is an important problem with potential complications such as contamination, infection and pneumothorax. Besides, mere loosening can also lead to malfunction. It is a common practice to tape the connection of the system. This study aimed to evaluate the materials and methods of connection of chest drain system to minimize drainage dislodgement.

METHODS:We conducted an experimental study to assess the tightness of the connection with various taping materials and methods. We selected three commonly used adhesive materials (3M?Transpore? Medical tape, 3M? Micropore? Medical tape, 3M? Soft Cloth Tape on Liner) and three different methods (cross method, straight method, nylon band) to secure the junction between the thoracostomy tube and the bi-conical adaptor in the drainage system. The measured outcome was the weight causing visible loosening of the junction between thoracotomy tube and the adaptor.

RESULTS:For each taping material and taping method, 10 trials were performed. The median weight required to disconnect the junction is 26.22 lb for Transpore?, 31.29 lb for Micropore? and 32.44 lb for Soft Cloth Tape on Liner. A smaller force was required to disconnect if Transpore? is used (P<0.001). There was no statistical signi fi cant difference between Micropore? and Soft Cloth Tape on Liner (P=0.98). The median disconnecting force is 32.44 lb for straight taping method, 40.55 lb for cross taping method and 21.15 lb for plastic band. The cross-taping method was the more secure method (P<0.0001 when compared with plastic band) (P=0.033 when compared with straight method).

CONCLUSION:Cross-taping is the most secure method among the tested varieties in connecting the thoracostomy tube to the chest drainage system. Transpore? is not a recommended material for thoracostomy tube taping.

Thoracostomy tube; Drainage system; Cross-taping

INTRODUCTION

Thoracostomy tube insertion is one of the common bedside procedures in emergency medicine and many acute specialties. According to the BTS guidelines for chest drain insertion, indications include pneumothorax, pleural effusion, empyema, traumatic hemopneumothorax and post-operative uses.[1]

Complications of tube thoracostomy can be classi fi ed into mechanical and infective.[2]Mechanical causes include tube malposition, blockage, dislodgement, re-expansion pulmonary edema, subcutaneous emphysema, nerve injuries, vascular injuries, esophageal injuries, residual pneumothorax, fistulae, tumor recurrence at insertion site, herniation through the site, chylothorax, and cardiac dysrhythmias. Infective complications include empyema and surgical site infection.

It is a common practice to tape the connection of the system after chest drain insertion and un-taped connection is rarely encountered. Tube dislodgement from the drainage system occurs occasionally, sometimes not identified immediately. It may not be the most common tube-related problem but it is associated with potential complications such as contamination, infection, persistent pneumothorax and misinterpretation of chest drain functioning. Such adverse effects can be minimized by proper taping and connection of the thoracotomy tube.

There are several common adhesive materials and methods in clinical use. However, there is no consensus on which one is better. Moreover, literature search has not identi fi ed any past study to assess this issue. This study evaluates the materials and methods of connection of the chest drain system to minimize drainage dislodgement.

METHODS

We conducted an experimental study to assess the tightness of the connection with various taping materials and methods.

Our study was divided into two parts.

In the first part, we selected three commonly used adhesive materials (3M? Transpore? Medical tape, 3M? Micropore? Medical tape, 3M? Soft Cloth Tape on Liner) to secure the junction ( a bi-conical-shaped adapter) between the thoracostomy tube and the drainage system (Figure 1). The straight method was used as this method is simple to apply and the application can be more standardized.

Figure 1. Three commonly used medical tape with adhesive materials. 1: 3M? Transpore? Medical tape; 2: 3M? Micropore? Medical tape; 3: 3M? Soft Cloth Tape on Liner.

Figure 2. Three different methods (straight method, cross method, plastic band) for junction connection. Straight method: Two longer tapes (A1 and A2, measured 2.5X9 cm) were applied longitudinally; another two shorter tapes (B1 and B2 measured 2.5X5 cm) were applied making circumferential turns; the connection junction (*) was not covered by the tapes. Cross (X) method: one single long tape measured 2.5X40 cm was used; one complete circumferential turn was made at one end; a diagonal turn was made; a complete circumferential turn was made at the other end; another diagonal turn was made, making an X-shape with step 2; completed with a fi nal circumferential turn; a "window" was created so that the connection junction (* ) was visible to detect any disconnection or subtle loosening.

In the second part, we used three different methods (straight method, cross method, plastic band) for junction connection (Figures 2 and 3). 3M? Soft Cloth Tape onLiner was used in the fi rst two methods for comparison. 3M? Soft Cloth Tape on Liner was chosen as it was commented to have strong and reliable adhesion in both dry and moist conditions according to the manufacturer information. It has durable backing resistance to tearing and stretching. It also has good adhesion to tubing. These features make it a desirable choice for thoracotomy connection.

In our study, Argyle 28 Fr straight thoracic catheters were used. The bi-conical adaptor was inserted into the catheter to the same depth, i.e. the third marking on the adaptor (measures 1.9 cm from the tip). It assumed similar intrinsic tightness of the connection before any taping material was applied. Thoracic catheters and connectors were reused while taping materials were newly prepared by ungloved hands.

Figure 3. Two plastic bands (3.6X30 cm, material: nylon 66) were applied at the two ends.

Figure 4. A thoracotomy system. A hook was fi xed on the wall. Fishing scale portable scale (Max weight: 110 lbs/Shenzhen West-Boao Science & Technology Co., Ltd). A nylon ring was applied through the end side hole of a 28 Fr straight Argyle chest catheter. A drainage tube was connected to the other end. Various strength of longitudinal force was generated by applying different weight to the weight reader (à). The measured outcome was the weight causing visible loosening of the junction (#) between the thoracotomy tube and the chest drainage system adapter.

A thoracotomy system was connected according to Figure 4. The measured outcome was the weight causing visible loosening of the junction between the thoracotomy tube and the chest drainage system adapter. It was measured by a weight reader which showed readings of 2 decimal places. The procedure was video-taped and then replayed. The maximal readings obtained until visible disconnection was taken as the disconnecting force.

Statistical analysis

The weight unit pound (lb) was used and taken to 2 decimal figures. The median values were rounded off to 2 decimal figures. The data were input in an Excel sheet and analyzed using a statistical package for social science (IBM SPSS version 20 for Windows).

The Kruskal-Wallis test was employed for comparison between multiple groups. Pairwise multiple comparisons were carried out with Bonferroni's correction. Level of signi fi cance was taken at 5% and P value of less than 0.05 was regarded as statistically signi fi cant.

RESULTS

For each taping material and taping method, ten trials were performed.

For part 1, the median weight required to disconnect the junction was 26.22 lb for Transpore?, 31.29 lb for Micropore? and 32.44 lb for Soft Cloth Tape on Liner (Table 1).

A smaller force was required to disconnect if Transpore? was used when compared with Micropore?and Soft Cloth Tape on Liner (P<0.0001). There was no statistical significant difference between Micropore?and Soft Cloth Tape on Liner (P=0.98) (Table 2).

Table 1. Median weight required to disconnect the junction in three different taping materials

Table 2. P-value of comparing three different taping materials

Table 3. Median weight required to disconnect the junction in three different connection methods

Table 4. P-value of comparing three different connection methods

For part 2, the median disconnecting force was 32.44 lb for straight taping method, 40.55 lb for cross taping method and 21.15 lb for nylon band (Table 3).

Cross-taping was the more secure method compared with the other two methods (P<0.0001 as compared with plastic band and P=0.033 as compared with straight method) (Table 4).

DISCUSSION

This study applied an objective assessment, instead of subjective feeling, for the strength of the various common methods of connection of the thoracostomy tube to the drainage system.

Micropore? is a latex-free, hypoallergenic paper tape with good adhesive power. Its advantage is gentle to skin. The suggested applications include small to medium dressing, damp skin, securing lightweight tubing, ostomy appliances, fragile skin and repeated taping.

Transpore? is another choice of latex-free and hypoallergenic tape material. It is easy to tear into very thin strips to tailor the tape for various dressings and devices. It is easy to use with gloves. The suggested applications are securing tubing and devices (e.g., catheters, IV tubing), bulky dressing and when tape width must be customized. One clinical use in our department is anchoring of an endotracheal tube on intubated patients. These features appeared to be advantageous to be used in our study. However, its connection strength was the least among the three from our study results.

Soft Cloth Tape on Liner is also hypoallergenic and latex-free. It provides strong adhesion to both skin and tubing. It has durable resistance against tearing and stretching. The "Liner" refers to the underlying removable paper with dimensions reference so that the width and length can be customized easily. It also adheres well in both moist and dry conditions. The suggested applications are securing tubing.

The cross method is recommended for two main reasons. The first reason is that it provides a strong connection. The second reason is that it creates a "window" making the connection easily visualized during nursing care. Part of the bi-conical adapter was not covered by the tapes to ensure proper fitting to the thoracostomy tube and drainage system. It can prevent non-visualization of the subtle loosening or disconnection.

The use of commercial nylon band is neat and simple but it may not be readily available in the clinical setting. Also, their strength is not strong enough.

There are some limitations in our study. First, the insertion depth between the bi-conical adaptor and thoracotomy catheter may not be identical, although same size and type of the catheters are used. And the depth is standardized by the same marking on the adaptor, but differences are still possible.

Second, the recommended cross method takes more technical steps than the straight method. It involves the cutting of adhesive tape into long strip and making circumferential turns. The tightness of applications may not be the same. However, most trained clinical staff could readily master the skills with some practice. Hence, the application can easily be generalized to the clinical settings.

Next, different sizes of thoracotomy tube are used in different clinical situations while 28 Fr straight Argyle chest catheters are used in our study. For instance, a larger bore is used in traumatic hemopneumothorax when a smaller one in spontaneous pneumothorax. While a small bore is used, it is dif fi cult to apply tapes by the cross method, as well as the recommended size of taping materials as stated in our study may not be applicable.

In conclusion, application of 3M Micropore? and Soft Cloth on Liner is more secure in connecting the thoracostomy tube to the chest drainage system. However, it is inconclusive on which one is more superior to another.

Cross-taping is the most secure method among the tested varieties in thoracotomy connection. It would be worthwhile to conduct another study on comparison between 3M Micropore? and Soft Cloth on Liner. A standardized use of taping material and method can prevent thoracotomy related complications and provide better care to our patients.

ACKNOWLEDGEMENT

The authors are greatly indebted to Mr Thomas Yeung and Mr SHEK Yi Nok for their substantial contribution in the conduction of the experiments of this study.

Funding:None.

Ethical approval:Not needed.

Con fl icts of interest:No work resembling the enclosed article has been published or is being submitted for publication elsewhere. We have disclosed all financial support for our work and other potential con fl icts of interests.

Contributors:All authors have each made a substantial contribution so as to qualify for authorship and that we have approved the contents.

REFERENCES

1 Laws D, Neville E, Duffy J; Pleural Diseases Group, Standards of Care Committee, British Thoracic Society. BTS guidelines for the insertion of a chest drain. Thorax 2003; 58 Suppl 2: ii53- ii59.

2 Kesieme EB, Dongo A, Ezemba N, Irekpita E, Jebbin N, Kesieme C. Tube thoracostomy: complications and its management. Pulm Med 2012; 2012: 256878.

Received April 6, 2014

Accepted after revision September 19, 2014

World J Emerg Med 2014;5(4):259–263

10.5847/wjem.j.issn.1920–8642.2014.04.003

主站蜘蛛池模板: 国产精品美女免费视频大全| 一级福利视频| 色婷婷狠狠干| 日韩精品一区二区深田咏美| 手机在线国产精品| 国产精品视频第一专区| 精品精品国产高清A毛片| 国产一区二区色淫影院| 一区二区无码在线视频| 蜜桃视频一区二区| 欧美激情视频二区三区| 美女国产在线| 欧美黄色网站在线看| 亚洲看片网| 伊人久久青草青青综合| 国产农村妇女精品一二区| 超清无码熟妇人妻AV在线绿巨人| 毛片免费视频| 国产99在线| 久久青草免费91线频观看不卡| 婷婷伊人久久| 天天做天天爱夜夜爽毛片毛片| 国产亚洲欧美在线中文bt天堂| 欧美v在线| 精品伊人久久久香线蕉| 黄色网页在线观看| 区国产精品搜索视频| 一级毛片免费观看久| 国产无码高清视频不卡| 黄色a一级视频| 亚洲综合色吧| 国产精品免费福利久久播放| 91毛片网| 国产成人一二三| 538精品在线观看| 色135综合网| 国产精品99久久久久久董美香| 免费午夜无码18禁无码影院| 99re在线免费视频| 免费高清毛片| 亚洲欧美日本国产综合在线| 中文字幕永久在线看| 素人激情视频福利| 国产成a人片在线播放| 午夜免费小视频| 亚洲视频免费播放| 国产在线精品美女观看| 国产高清精品在线91| 性视频久久| 99ri国产在线| 九九九精品视频| 日韩精品一区二区三区swag| 亚洲欧洲日本在线| 国产办公室秘书无码精品| 国产三级成人| 亚洲美女一级毛片| 91激情视频| 天堂久久久久久中文字幕| 国产成人高清精品免费| 午夜国产在线观看| 亚洲国产看片基地久久1024| 69视频国产| 国产va在线| 精品国产美女福到在线不卡f| 久996视频精品免费观看| 青青青国产免费线在| 欧美一级在线| 国产亚洲精品yxsp| 一级毛片免费的| 青青草原国产| 97se综合| 丁香五月激情图片| 中文字幕 欧美日韩| 欧美一级特黄aaaaaa在线看片| 91成人免费观看在线观看| 亚洲自拍另类| 国产精品视频第一专区| 91久久夜色精品| 自偷自拍三级全三级视频 | 激情视频综合网| 98精品全国免费观看视频| 精品国产免费观看|