FITTING CORRECTION METHOD OF RING ARTIFACTS FOR RECONSTRUCTING CONE-BEAM CT IMAGES

2010-10-08 12:10:24LuoShouhuaWuJingZhangBoChenGong

Luo Shouhua,Wu Jing,Zhang Bo,Chen Gong

(1.School of Biological Science&Medical Engineering,Southeast University,Nanjing,210096,P.R.China;2.Information Engineering Department,Jiangsu Provincial Hospital of TCM,Nanjing,210029,P.R.China)

INTRODUCTION

Ring artifacts on cone-beam computed tomography(CBCT)images are introduced by the nonuniformity of the flat-panel detector element response,which always appear as rings with different radius centered on the rotational axis.It is common in the process of CT image reconstruction. The reasons for such unsatisfactory responses are as follows:pixel defects of the detector;pixel nonlinear response on various X-ray intensity;inconsistent responses among pixels on certain X-ray intensity;pixel nonlinear responses on various X-ray spectra;instability of dark-field and bright-field of thedetector,etc[1].Due to the reason that data of the constructed images are corrupted by these ring artifacts,qualitative and quantitative analyses of CBCT images will be compromised.

Presently,there are many methods to reduce ring artifacts,including hardware and software approaches. The hardware approach is mainly used to improve the system scan sampling procedures[2],thus pixels with inconsistent response can obtain different radial signals,and then all inconsistencies of the detector systematic response will be uniformly distributed over whole sampling sequences on the initial projection and acquisition stages,resulting in effective reductions of ring artifacts.The software approach includes filtering and correcting methods.The filtering method is a post-processing technique in the image reconstruction.Aimed at the characteristics of ring artifact on the CT image,the researchers make a number of target filters,which can filter out artifacts and reserve the high-frequency edges of the images at the same time[3-5].However,these filters increase the reconstruction timeand result in blurs and degradation of the reconstruction images.The correcting method can overcome these two shortcomings by preprocessing the projective images.Different correcting methods depend on different factors of ring artifacts[6-7].Theclassical correcting methods only deal with the offset errors,and such kind of correction does not completely suppress the inconsistency of the pixel re-sponse.This paper uses the fitting correction method to suppress the ring artifacts of theimage reconstruction,which makes the curves of different pixels response to the radial intensity and suppress the ring artifacts.Finally,how to improve the fitting correction is discussed.

1 MATERIALSAND METHODS

1.1 Ring artifact f ormation

As an example,for simplicity,two-dimensional parallel beam projection is introduced.As shown in Fig.1,sot coordinate system is presented by the xoy coordinate system with rotatingθ angle around theorigin o.One-dimensional detector is located inside s axis and parallel-beam rays parallel to t axis.θis the angleof the ray-detector system rotating around body;f(x,y)the point on the projected plane and f(r,h)the corresponding point in polar coordinates against the xoy coordinate system.

Fig.1 Geometric diagram of ring artifact caused by inconsistency of channel response

According to the geometric relationship,the projection position of f(x,y)located on s axis is

The ray-detector system rotates a circle around the rotation axis,that is,the angleθincreases from 0°to 360°.If r is invariable and h keeps synchronous increase withθ,the value of s will be constant.That is,the projection location of all points on the round circle with thecenter at the origin and the radius r on projection plane is the sameon thedetector.If responses of different pixels are inconsistent,the data of every round circle will not alter smoothly,and ring artifacts occur.

1.2 Methods

Before correcting projection images,the response curve of the pixel to the X-ray intensity of Hamamatsu C7921CA-02 detector is shown in Fig.2.

Fig.2 Pixel curves of responses vs.tube current(photon flux)

As seen in Fig.2,the response curve of the pixel to the intensity ray is nearly linear.However,the slope and the intercept of curves vary in corresponding response curves which make the response inconsistent among pixels.If all response curves tend to be the average one,as shown in Fig.2(b),the response inconsistency can be suppressed.

where i is the current output intensity of the X-ray source;Gi(x,y)the output gray value of the pixel(x,y).a(x,y)and b(x,y)are the slope and the intercept of the response curve of pixel(x,y);andthe slope and the intercept of all pixel average response curve,respectively,thus

In the fitting process,the current variable i would not be allowed because of different dimensionalities with the image gray value.Through Eqs.(2-3),the current variables can be eliminated.

Assuming that

Eq.(4)is simplified as

where c(x,y)and d(x,y)are the correction factors that need to be solved with the fitting correction method.Through Eq.(7),correction factors of each pixel can be solved by fitting a straight line.The details of whole procedure are as follows:capture n pieces of idle load pictures which are exposed by X-ray with same voltage strength and different currentintensities.Theaverageoutput gray scale of each picture is set as x-coordinate. And y-coordinate represents the output gray scale value of each pixel(x,y)exposed by different current intensities. And then, the straight lineis fitted across these points by a least square algorithm.As a result,the scope and the intercept of the line represent correction factors c(x,y)and d(x,y).Finally,the projected image is corrected through Eq.(8)

G′i(x,y)=(Gi(x,y)-d(x,y))/c(x,y) (8)where Gi(x,y)is the gray scale value of the uncorrected projected image and G′i(x,y)the gray scale value of the corrected projected image.

2 RESULTS OF RECONSTRUCTION IMAGE

The results of the reconstructed images by the fitting correction of the projected images in micro-CT are shown in Fig.3.

Fig.3 Reconstruction images of foam aluminum materials

In order to achieve a good suppressive effect,more details and improvements for thefitting correction method are discussed:

(1)The variability of the correction factors c(x,y)and d(x,y)is presented as follows:when the idle load pictures are obtained with the same current intensity and different current-voltages under the X-ray irradiation,the fitting correction results are shown in Fig.4 with different frame numbers of 1 and 16.Fig.4(a)shows the idle load picture taken in one frame,Fig.4(b)shows the idle load picture taken up to 16 frames.From the picture,the ring artifact removal efficiency of Fig.4(b)is better than that of Fig.4(a).This is because the ups and downs of ray photons will cause a certain degree of inconsistency,and taking the average of more frames can help to eliminate this inconsistency.While the frame count is up to a certain extent,the effect on the reconstruction does not promote any more.In this paper,16 frames of each angle should be enough to meet needs for removing the ring artifacts.

Fig.4 Comparison of fitting results with different idle load picture frames

(2)Conventionally,the correction factor of each pixel is statically treated,but actually,there will be some bias.On the same X-ray voltage(70 k V)and different current intensities(30μA and 45μA),we obtain the idle load pictures of 400 angles,which are performed by fitting correction reconstruction with the same correction factor.The partial enlarging pictures of the same slice are show n in Fig.5.In Fig.5,Fig.5(a)is the reconstruction result of 30μA current intensity;Fig.5(b)is the reconstruction result of 45μA current intensity.It is obvious that the ring artifact of Fig.5(b)is clearer than that of Fig.5(a).The result shows that the same correction factor does not produce thesame result exactly when the same pixel of detector is radiated by different X-ray intensities.The fitting lines of correction factors are nonlinear.How to obtain these specific nonlinear factors is another improvement aspect.Fig.5 shows reconstructed idle load images exposed in X-ray sources with samevoltagestrength and different current intensities.

(3)As correction factors are solved by fitting correction,more fitting points are better for the result,especially for fitting points nearby the ring artifact areas.

Fig.5 Partial enlarged pictures of same slice

3 CONCLUSION

This paper uses the fitting correction method to reduce the ring artifacts of micro-CT reconstruction images.The fitting correction preprocesses the projection images and makes the entire pixel response curve linear so as to suppress the ring artifacts on the reconstructed images.The results testify the effectiveness of the method.More improvement aspects of the fitting correction are discussed,such as the average of multiframeimages and increasing fitting points,etc.

In addition,this paper mainly deals with ring artifacts caused by inconsistencies of the pixel response.There are also many other factors leading to ring artifacts,such as nonlinear factors in fitting straight line for correction factors which can be targeted on theimplementation in the future.References:

[1] Li Junjiang,Hu Shaoxing,Li Baolei,et al.Ring artifact correction for industrial CT images[J].Journal of Beijing University of Aeronautics and Astronautics,2007,33(11):1378-1382.(in Chinese)

[2] Fu Jian,Lu Hongnian.Correcting method for ring

artifacts in fan-beam X-ray ICT[J].Optics and Preci-sion Engineering,2002,10(6):542-546.(in Chinese)

[3] Yang Chen,Ma Jianhua,Feng Qianjin,et al.Nonlocal prior Bayesian tomographic reconstruction[J].Math Imaging,2008,30(2):133-146.

[4] Yang Jun,Zhen Xin,Zhou Linghong,et al.Geometric correction for cone-beam CT reconstruction and artifacts reduction[C]//The 2nd International Conference on Bioinformatics and Biomedical Engineering.Shanghai,China:IEEE,2008:2386-2389.

[5] Sun Haining,Qiu Shaokun,Lou Shanshan,et al.A correction method for nonlinear artifacts in CT Imaging[C]//Proceedings of the26th Annual International Conference of the IEEE.San Francisco:IEEE,2004:1290-1293.

[6] Zhou Zhenggan,Teng Shenghua,Jiang Wei,et al.X-ray flat-panel-detector-based digital radiography and its image calibration[J].Journal of Beijing University of Aeronautics and Astronautics,2004,30(8):698-701.(in Chinese)

[7] Li Junjiang,Lu Hongnian,Li Baolei.Study on nonuniformity correction of X RII-CCD pixel channel response based on BP neural network[J].Optical Technique,2007,33(2):273-280.(in Chinese)

主站蜘蛛池模板: 无码日韩视频| www亚洲天堂| 91香蕉视频下载网站| 欧美在线精品怡红院| 女人18毛片水真多国产| 亚洲高清国产拍精品26u| 无码 在线 在线| 日韩精品一区二区三区大桥未久| 青青热久麻豆精品视频在线观看| 成人午夜免费视频| 黄色网在线| 69免费在线视频| 国产精品手机在线播放| 中文国产成人精品久久一| 色综合成人| 久久不卡国产精品无码| 精品无码日韩国产不卡av| 香蕉国产精品视频| 国产精品久线在线观看| 四虎在线观看视频高清无码| 国产黑丝一区| 毛片网站在线看| 超薄丝袜足j国产在线视频| 真人免费一级毛片一区二区| 浮力影院国产第一页| 精品人妻一区无码视频| 伊人久久青草青青综合| 久久国产高清视频| 国产在线麻豆波多野结衣 | 91久久性奴调教国产免费| 亚洲精品免费网站| 国产精品精品视频| 亚洲第一成年人网站| 亚洲午夜国产精品无卡| 在线视频亚洲色图| 极品av一区二区| 久久久噜噜噜久久中文字幕色伊伊 | 欧美国产综合视频| 国产精品丝袜视频| 丁香六月激情综合| 精品午夜国产福利观看| 久久黄色视频影| 99久久精品久久久久久婷婷| 日本黄色不卡视频| 久久婷婷人人澡人人爱91| 91视频青青草| 亚洲av无码专区久久蜜芽| 91在线一9|永久视频在线| 欧美精品xx| 久久人妻xunleige无码| 无码内射在线| 国产网站在线看| 日韩高清在线观看不卡一区二区| 久久综合伊人77777| 强奷白丝美女在线观看| 亚洲手机在线| 国产成人精品无码一区二| 天堂成人av| 国产激情无码一区二区免费| 成人91在线| 五月天丁香婷婷综合久久| 国产成人AV综合久久| 中文字幕久久波多野结衣| 亚洲第一区欧美国产综合| 99热最新网址| 最近最新中文字幕免费的一页| 亚洲Av激情网五月天| 中文国产成人久久精品小说| 精品久久国产综合精麻豆| 国产精品无码一二三视频| 久久综合AV免费观看| 亚洲欧美精品一中文字幕| 久久精品中文字幕少妇| 精品一区二区三区水蜜桃| 中国精品自拍| 国产色网站| 久996视频精品免费观看| 日韩精品资源| 97免费在线观看视频| 婷婷亚洲最大| 亚洲第一区精品日韩在线播放| 操美女免费网站|